What is Parallel Accumulation and Serial Fragmentation (PASEF)?

Parallel accumulation-serial fragmentation (PASEF) is a spectroscopic technique used during liquid chromatography-mass spectrometry (LC-MS) based proteomics to improve sequencing speed and sensitivity.

Structure of a protein - to be analysed using PASEF - By molekuul_bemolekuul_be | Shutterstock

How does PASEF work?

PASEF uses trapped ion mobility spectrometry (TIMS) to accumulate precursors released in a particular order depending on the mobility of the ion. Following separation by liquid chromatography, precursors accumulate in a tunnel where the opposing forces of gas-flow and an electric field hold them in place.

These precursors find an equilibrium between these forces at a particular point within the tunnel as a function of their collisional cross section, as a larger cross-section leads to a greater number of collisions and thus larger force applied by colliding gas molecules.

Once the desired number of ions have accumulated, they are eluted by slowly lowering the strength of the magnetic field, entering the mass spectrometry device to undergo fragmentation.

What are the advantages of PASEF?

As the precursor ions accumulate depending on their mass and charge before eluting into the mass spectrometer, the signal-to-noise ratio for each precursor species is massively improved due to the larger concentration of ions eluting simultaneously.

Further separation between precursor ions of the same mass-to-charge ratio due to different ion mobilities is possible. For example, ions of different species with the same mass and charge may have differing collisional cross sections, affecting their order of elution. This alleviates ‘precursor ion fraction problem’, which occurs when species in low abundance are occluded by more abundant precursor fractions of the same mass-to-charge ratio.

Traditionally, only a single precursor can be selected for fragmentation at a time, while PASEF allows multiple precursors to be fragmented during a single scan, without loss of signal or sensitivity. Additionally, resultant spectra are fully mass-resolved with the additional factor of ion mobility considered.

These result in an approximately ten-fold increase in shotgun proteomics sequencing speed, due to the accumulation of precursor ions in parallel, leading to an improved signal-to-noise ratio that overcomes the diminishing returns of increasingly fast mass spectrometry mass spectra acquisition.

The increased sensitivity denoted by PASEF also allows for smaller volumes of sample to be consumed in the process, making it particularly attractive for applications with limited quantities of a sample available, such as while performing a tumor biopsy.

Challenges of LC-MS that are overcome with PASEF

In regular LC-MS for proteomics, a protein is usually digested using enzymes and the resulting peptide mixture is separated by liquid chromatography. The separated components (precursor ions) of the peptide mixture are then electrosprayed into a mass spectrometer, where particular precursors are isolated depending on their mass-to-charge ratio (m/z) and fragmented through collisions with ions or electrons.

The precursors are selected by a survey spectrum, followed by fragmentation of the most abundant precursors only. However, as little as 16% of precursors generated by liquid chromatography are fragmented during mass spectrometry, as multiple precursors often elute from the liquid chromatography column at the same time, lowering the acquisition time available for each precursor. This leads to many potentially insightful precursors being discarded before fragmenting.

Further Reading

Last Updated: Jan 24, 2019

Michael Greenwood

Written by

Michael Greenwood

Michael graduated from the University of Salford with a Ph.D. in Biochemistry in 2023, and has keen research interests towards nanotechnology and its application to biological systems. Michael has written on a wide range of science communication and news topics within the life sciences and related fields since 2019, and engages extensively with current developments in journal publications.  

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Greenwood, Michael. (2019, January 24). What is Parallel Accumulation and Serial Fragmentation (PASEF)?. News-Medical. Retrieved on January 02, 2024 from https://www.news-medical.net/life-sciences/What-is-Parallel-Accumulation-and-Serial-Fragmentation-(PASEF).aspx.

  • MLA

    Greenwood, Michael. "What is Parallel Accumulation and Serial Fragmentation (PASEF)?". News-Medical. 02 January 2024. <https://www.news-medical.net/life-sciences/What-is-Parallel-Accumulation-and-Serial-Fragmentation-(PASEF).aspx>.

  • Chicago

    Greenwood, Michael. "What is Parallel Accumulation and Serial Fragmentation (PASEF)?". News-Medical. https://www.news-medical.net/life-sciences/What-is-Parallel-Accumulation-and-Serial-Fragmentation-(PASEF).aspx. (accessed January 02, 2024).

  • Harvard

    Greenwood, Michael. 2019. What is Parallel Accumulation and Serial Fragmentation (PASEF)?. News-Medical, viewed 02 January 2024, https://www.news-medical.net/life-sciences/What-is-Parallel-Accumulation-and-Serial-Fragmentation-(PASEF).aspx.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of News Medical.
Post a new comment
Post