Flex Pipettes are designed for automation and provide incredibly accurate pipetting from 1 to 96 channels.
Accurate from 1 to 1000 µL
Image Credit: Opentrons
Thanks to air displacement technology, flexible pipettes provide extremely accurate pipetting over large volume ranges.
Modular pipetting heads
Image Credit: Opentrons
Users can quickly and easily install or replace their pipettes independently in a few minutes.
Automated calibration
Video Credit: Opentrons
Smart sensors provide automatic calibration, real-time positioning, and detection of errors.
Tips built for automation
Image Credit: Opentrons
Tips tailored for automation and created especially for the Flex.
Tech specs
Source: Opentrons
Pipette |
Volume |
Random Error |
Systematic Error |
|
µL |
%D |
µL |
%CV |
µL |
Flex 1-Channel Pipette (1–50 µL) |
1
10
50 |
±8.00%
±1.50%
±1.25% |
0.08 µL
0.15 µL
0.625 µL |
±7.00%
±0.50%
±0.40% |
0.07µl
0.05µl
0.20µl |
Flex 1-Channel Pipette (5–1000 µL) |
5
50
200
1000 |
±5.00%
±0.50%
±0.50%
±0.50% |
0.25 µL
0.25 µL
1.00 µL
5.00 µL |
±2.50%
±0.30%
±0.15%
±0.15% |
0.125 µL
0.15 µL
0.30 µL
1.50 µL |